Effect of biotin on currently used Beckman thyroglobulin assay and newly reformulated thyroglobulin assay not affected by biotin

Bailey Countryman,Melissa McBride,Trevor McCracken,Amitava Dasgupta
DOI: https://doi.org/10.1093/ajcp/aqae048
2024-04-29
American Journal of Clinical Pathology
Abstract:Abstract Objectives Biotin causes negative interference with thyroglobulin measurement using the Access thyroglobulin assay. Recently, Beckman reformulated the thyroglobulin assay to overcome biotin interference. We investigated the effect of biotin on both current and newly formulated assays. Methods Four serum pools were prepared using specimens containing various amounts of thyroglobulin. Then aliquots of each pool were supplemented with various amounts of biotin, and thyroglobulin concentrations were measured by both the current and the new assays. In addition, 3 volunteers ingested 10 mg biotin, and specimens were drawn before and 2 hours after taking biotin. Thyroglobulin concentrations before and 2 hours after taking biotin were measured by both assays. Results In the presence of biotin, thyroglobulin concentrations were reduced significantly using the current assay, but no significant change was observed using the newly formulated assay. We observed similar results in vivo. Conclusions The newly formulated thyroglobulin assay by Beckman is free from biotin interference.
pathology
What problem does this paper attempt to address?