Non-Destructive Sensor for Glucose Solution Concentration Detection Using Electromagnetic Technology

Shasha Yang,Shiwen Gao,Yi Zhuang,Wence Hu,Junyi Zhao,Zhenxiang Yi
DOI: https://doi.org/10.3390/mi15060758
IF: 3.4
2024-06-06
Micromachines
Abstract:In this paper, a sensor using a complementary split ring resonator (CSRR) is proposed for non-destructive testing of blood glucose. By depicting the complementary split ring structure on the ground, the electromagnetic field strength between the split rings can be enhanced effectively. The structure size of the sensor by CSRR is determined by simulation, so that the insertion loss curve of the device has a resonance point at the frequency of 3.419 GHz. With a special holder created by three-dimensional (3D) printing technology, the test platform was established when the concentration of the solution varied from 0 mg/mL to 20 mg/mL. The experimental results indicate that there is an obvious linear relationship between the insertion loss S21 and the glucose concentration at the resonant frequency. Similarly, the measured real part and imaginary part of the S21 both vary with glucose concentration linearly. Based on the above experimental results, the feasibility of the sensor using a CSRR proposed in this paper for non-destructive detection of blood glucose is preliminarily verified.
chemistry, analytical,nanoscience & nanotechnology,instruments & instrumentation,physics, applied
What problem does this paper attempt to address?
The problem this paper attempts to address is the development of a non-invasive glucose concentration detection sensor. Specifically, the authors propose a sensor based on Complementary Split Ring Resonators (CSRR) for non-destructive detection of glucose concentration in blood. Traditional glucose detection methods usually require blood sampling, which not only imposes physiological and psychological burdens on patients but also increases the risk of infection. Therefore, developing a non-invasive, real-time glucose monitoring technology is of great significance for the prevention and clinical treatment of diabetes. The main contributions of the paper include: 1. **Sensor Design**: By depicting the complementary split ring structure on the ground plane, the electromagnetic field intensity between the split rings is enhanced, thereby improving the sensitivity of the sensor. 2. **Experimental Validation**: A test platform was established using a solution holder made with 3D printing technology, and experiments were conducted in glucose solutions of different concentrations to verify the feasibility of the sensor. 3. **Linear Relationship**: Experimental results show that at the resonant frequency, there is a significant linear relationship between the insertion loss (S21) and the glucose concentration, providing a theoretical basis for non-invasive glucose detection. In summary, this study aims to develop an efficient, non-invasive glucose detection method to reduce the burden on patients and improve the convenience and accuracy of diabetes management.