Performance evaluation of an automatically controlled transcutaneous energy transfer system powering left ventricular assist device

Sarath S. Nair,D. S. Nagesh,Sachin J. Shenoy,S. Harikrishnan
DOI: https://doi.org/10.1111/aor.14892
2024-10-24
Artificial Organs
Abstract:Wireless powering through a Transcutaneous Energy Transfer System (TETS) is a promising solution for eliminating infections in drive line of LVAD. Automatically Controlled TETS is developed to compensate for the variations in efficiency and power transfer during dynamic clinical conditions. In vitro and animal models performance and stability evaluation confirms the potential benefit of automatically controlled TETS for powering a Left Ventricular Assist Device. Power up to 20 W could be transferred at an axial gap of 30 mm. The LVAD could be operated at the required RPM irrespective of the change in axial positions. A 5% variation in speed and a 2% change in flow rate are observed with the change in tissue thickness, in porcine carcass model. Background The survival rate of LVAD recipients may be improved by eliminating the infection failure mode at the percutaneous lead entry site. Wireless powering through a Transcutaneous Energy Transfer System (TETS) is a promising solution for achieving this. However, automatic controls need to be employed to compensate for the variations in efficiency and power transfer due to changes in load and coil‐to‐coil gaps. Methods This article discusses the results of in vitro and animal models performance evaluation of an automatically controlled TETS for powering a Left Ventricular Assist Device. Results It was found that power up to 20 W could be transferred at an axial gap of 30 mm. A 5% variation in speed and a 2% change in flow rate are observed with the change in tissue thickness, in porcine carcass model. Conclusions The LVAD could be operated at the required RPM irrespective of the change in axial positions.
engineering, biomedical,transplantation
What problem does this paper attempt to address?