A simple step to improve fat and energy delivery of human milk delivered via bottle‐feeding pump, an experimental study

Caitlin Irwin,Kate Beard,Louis Martin,Kok Lim Kua,Katie Huff
DOI: https://doi.org/10.1002/jpen.2659
2024-06-19
Journal of Parenteral and Enteral Nutrition
Abstract:Background Enteral feeding pump systems deliver decreased amounts of macronutrients in human milk to neonates. This study determined the macronutrient loss associated with a bottle‐feeding pump system and the effect of manually mixing the human milk during extended feeds. Methods Macronutrient content from samples of donor human milk was analyzed after simulated extended feeds with a bottle‐feeding pump system, using a human milk analyzer. Simulations were repeated using manual mixing of the bottle every 30 min during feeding. The percentage of the baseline was calculated, and one‐sample t tests and analysis of variance compared the effect of manual mixing and the duration of feeding on macronutrient delivery. Results The delivery of fat and energy was lower over time, but manual mixing considerably improved retention. The length of feeding impacted fat delivery, with less fat delivered over time (P