Energy Calibration of the JLab Bremsstrahlung Tagging System

S. Stepanyan,S. Boyarinov,H. Egiyan,L. Guo,D. Dale,M. Gabrielyan,L. Gan,A. Gasparian,A. Glamazdin,B. Mecking,I. Nakagawa,A. Teymurazyan,M.H. Wood
DOI: https://doi.org/10.1016/j.nima.2006.12.009
2006-12-20
Abstract:In this report, we present the energy calibration of the Hall B bremsstrahlung tagging system at the Thomas Jefferson National Accelerator Facility. The calibration was performed using a magnetic pair spectrometer. The tagged photon energy spectrum was measured in coincidence with $e^+e^-$ pairs as a function of the pair spectrometer magnetic field. Taking advantage of the internal linearity of the pair spectrometer, the energy of the tagging system was calibrated at the level of $\pm 0.1% E_\gamma$. The absolute energy scale was determined using the $e^+e^-$ rate measurements close to the end-point of the photon spectrum. The energy variations across the full tagging range were found to be $<3$ MeV.