Development of the CE-C 4 D method for determination of organic acids in infants faeces using an in-house built instrument

Marcelina Rusin,Joanna Pluta,Aneta Woźniakiewicz,Justyna Dobrowolska-Iwanek,Hubert Huras,Magdalena Staśkiewicz,Ryszard Lauterbach,Michał Woźniakiewicz
DOI: https://doi.org/10.1016/j.microc.2024.110203
IF: 5.304
2024-02-25
Microchemical Journal
Abstract:Short-chain fatty acids (SCFAs) are a major group of products produced during colonic bacterial fermentation of nondigestible carbohydrates demonstrating several important functions, e.g. regulating the immune system activity or the lipid metabolism. A detailed profile and concentration characterization of SCFAs in infant faeces is essential for expanding knowledge regarding the role of SCFAs in infant physiology and pathophysiology. This research presents the development and validation of a capillary electrophoresis method with a contactless capacitively coupled conductivity detector (CE-C 4 D) for the analysis of SCFAs and succinic and lactic acids in infant faeces. The study focused on optimizing the background electrolyte, mainly dynamic capillary coating with polybrene, and utilizing cyclodextrins to achieve efficient separation. Validation parameters, such as linearity (0.993 – 0.998), limits of detection and quantification (LOD: 0.10 – 0.89 μg/mL; LOQ: 0.35 – 2.97 μg/mL), intra- and inter-day precision (1.71 – 5.69 % and 1.11 – 12.5 %, respectively), injection repeatability (0.91 – 5.70 %), trueness (-12.8 – 3.80 %), and matrix effect (78.8 – 108.6 %), were assessed and the method was found suitable for the purpose. The demonstrated method was successfully applied to determine nine organic acids in faecal samples.
chemistry, analytical
What problem does this paper attempt to address?