Performance of an enhanced afterloader with electromagnetic tracking capabilities for channel reconstruction and error detection

Daline Tho,Marie‐Claude Lavallée,Luc Beaulieu
DOI: https://doi.org/10.1002/mp.14877
IF: 4.506
2021-07-17
Medical Physics
Abstract:<section class="article-section__content"><h3 class="article-section__sub-title section1"> Purpose</h3><p>To assess catheter reconstruction and error detection performance of an afterloader (Elekta Brachytherapy, Veenendaal, The Netherlands) equipped with electromagnetic (EM) tracking capabilities.</p></section><section class="article-section__content"><h3 class="article-section__sub-title section1"> Materials/Methods</h3><p>A Flexitron research unit equipped with a special check‐cable in which is integrated an EM sensor (NDI Aurora V3) that enables tracking and reconstruction capability. Reconstructions of a 24 cm long catheter were performed using two methods: continuous fixed‐speed check cable backward stepping ( at 1, 2.5, 5, 10, 25 and 50 cm/s) and stepping through each dwell position every 1 mm. The ability of the system to differentiate between two closely located (parallel) catheters was investigated by connecting catheters to the afterloader and moving it from its axis with an increment of 1 mm. A robotic arm (Meca500, Mecademic, Montreal) with an accuracy of 0.01 mm was used to move the catheter between each reconstruction. Reconstructions were obtained with a locally weighted scatter‐plot smoothing algorithm. To quantify the reconstruction accuracy, distances between two catheters were computed along the reconstruction track with a 5 mm step. Reconstructions of curve catheter paths were assessed through parallel and perpendicular phantom configuration to the EM field generator. Indexer length and lateral error were simulated and ROC analysis was made.</p></section><section class="article-section__content"><h3 class="article-section__sub-title section1"> Results</h3><p>Using a 50 cm/s check cable speed does not allow accurate reconstruction. A slower check cable speed results in better reconstruction performance and smaller standard deviations. At 1 cm/s, a catheter can be shifted laterally down to a 1 mm and all paths can be uniquely identified. The optimum operating distance from the field generator (50 to 300 mm) resulted in a lower absolute mean deviation from the expected value (0.2±0.1mm) versus being positioned on the edge of the electromagnetic sensitive detection volume (0.6±0.3mm). Reconstructions of curved catheters with a check cable speed under 5 cm/s gave a 0.8 mm±0.3 mm error, or better. All indexer and lateral shifts of 1 mm were detected with a check cable speed of 2.5 cm/s or under.</p></section><section class="article-section__content"><h3 class="article-section__sub-title section1"> Conclusions</h3><p>The EM‐equipped Flexitron afterloader is able to track and reconstruct catheters with high accuracy. A speed under 5 cm/s is recommended for straight and curved catheter reconstructions. It allows catheter identification down to 1 mm inter‐catheter distance shift. The check cable can also be used to detect common shift errors.</p></section>
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?