100 ps time-of-flight resolution of Dielectric Resistive Plate Chamber

A. Akindinov,V. Golovin,A. Martemiyanov,V. Petrov,V. Plotnikov,A. Smirnitsky,K. Voloshin
DOI: https://doi.org/10.48550/arXiv.physics/0502111
2005-02-21
Abstract:Time of flight of a minimum ionizing particle along a fixed base has been measured with a 100 ps accuracy by means of a Dielectric Resistive Plate Chamber (DRPC) with 4 x 0.3 mm gas gaps. DRPC timing characteristics have been studied with different applied voltages, discriminating thresholds and beam intensities. It may be stated that the time-of-flight resolution of gaseous detectors developed within the ALICE experiment has reached the level of the best known scintillation counters.
Instrumentation and Detectors