Measurement of 222Rn dissolved in water at the Sudbury Neutrino Observatory

I. Blevis
DOI: https://doi.org/10.1016/j.nima.2003.10.103
2003-12-11
Abstract:The technique used at the Sudbury Neutrino Observatory (SNO) to measure the concentration of 222Rn in water is described. Water from the SNO detector is passed through a vacuum degasser (in the light water system) or a membrane contact degasser (in the heavy water system) where dissolved gases, including radon, are liberated. The degasser is connected to a vacuum system which collects the radon on a cold trap and removes most other gases, such as water vapor and nitrogen. After roughly 0.5 tonnes of H2O or 6 tonnes of D2O have been sampled, the accumulated radon is transferred to a Lucas cell. The cell is mounted on a photomultiplier tube which detects the alpha particles from the decay of 222Rn and its daughters. The overall degassing and concentration efficiency is about 38% and the single-alpha counting efficiency is approximately 75%. The sensitivity of the radon assay system for D2O is equivalent to ~3 E(-15) g U/g water. The radon concentration in both the H2O and D2O is sufficiently low that the rate of background events from U-chain elements is a small fraction of the interaction rate of solar neutrinos by the neutral current reaction.
Nuclear Experiment