Development of a large-mass, low-threshold detector system with simultaneous measurements of athermal phonons and scintillation light
M. Chaudhuri,G. Agnolet,V. Iyer,V.K.S. Kashyap,M. Lee,R. Mahapatra,S. Maludze,N. Mirabolfathi,B. Mohanty,M. Platt,A. Upadhyay,S. Sahoo,S. Verma
DOI: https://doi.org/10.1016/j.nima.2023.168374
2023-05-20
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Abstract:We have combined two low-threshold detector technologies to develop a large-mass, low-threshold detector system that simultaneously measures the athermal phonons in a sapphire detector while an adjacent silicon high-voltage detector detects the scintillation light from the sapphire detector. This detector system could provide event-by-event discrimination between electron and nuclear events due to the difference in their scintillation light yield. While such systems with simultaneous phonon and light detection have been demonstrated earlier with smaller detectors, our system is designed to provide a large detector mass with high amplification for the limited scintillation light. Future work will focus on at least an order of magnitude improvement in the light collection efficiency by having a highly reflective detector housing and custom phonon mask design to maximize light collection by the silicon high-voltage detector.
physics, particles & fields, nuclear,nuclear science & technology,instruments & instrumentation