Comparison of time taken to obtain an arterial blood gas result at the bedside using the ProximaTM point of care machine vs. a standard remote arterial blood gas analyser: A randomized controlled trial

G. Troughton,K. Mitchell,David Egbosimba,K. Salmon,M. Grocott
DOI: https://doi.org/10.1177/1751143720973847
2020-12-06
Journal of the Intensive Care Society
Abstract:Introduction The ProximaTM point of care (POC) device enables arterial blood gas (ABG) samples to be analysed without the nurse leaving the patient. The benefits of this for work efficiency have not been evaluated. Methods We compared the time taken to obtain an ABG result using ProximaTM versus a standard ABG sampling system. Twenty patients were randomized to ABG sampling using ProximaTM, or a standard ABG system. Nurses were observed performing all ABG sampling episodes for a minimum of 24 hours and no more than 72 hours. Results The mean time taken to obtain a result using ProximaTM was 4:56 (SD = 1:40) minutes compared to 6:31 (SD = 1:53) minutes for the standard ABG technique (p < 0.001). Mean time away from the patient's bedside was 3.07 (SD = 1:17) minutes using the standard system and 0 minutes using ProximaTM (p < 0.001). Conclusions Reduced time for blood gas sampling and avoidance of time away from patients may have significant patient safety and resource management implications, but the clinical and financial significance were not evaluated.
Medicine
What problem does this paper attempt to address?