Beam Diagnostics Systems For The National Ignition Facility

R. D. Demaret,R. D. Boyd,E. S. Bliss,A. J. Gates,J. R. Severyn
DOI: https://doi.org/10.48550/arXiv.physics/0111049
2001-11-09
Abstract:The National Ignition Facility laser focuses 1.8 Mega-joules of ultraviolet light (wavelength 351 nano-meters) from 192 beams into a 600-micro-meter-diameter volume. Effective use of this output in target experiments requires that the power output from all the beams match within 8% over their entire 20-nanosecond waveform. The scope of NIF beam diagnostics systems necessary to accomplish this task is unprecedented for laser facilities. Each beam line contains 110 major optical components distributed over a 510 meter path, and diagnostic tolerances for beam measurement are demanding. Total laser pulse energy is measured with 2.8% precision, and the inter-beam temporal variation of pulse power is measured with 4% precision. These measurement goals are achieved through use of approximately 160 sensor packages that measure the energy at five locations and power at 3 locations along each beamline using 335 photodiodes, 215 calorimeters and 36 digitizers. Successful operation of such a system requires a high level of automation of the widely distributed sensors. Computer control systems provide the basis for operating the shot diagnostics with repeatable accuracy, assisted by operators who oversee system activities and setup, respond to performance exceptions, and complete calibration and maintenance tasks.
Instrumentation and Detectors
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to ensure that the 192 lasers in the National Ignition Facility (NIF) in the United States can output energy precisely and consistently in target experiments. Specifically, the NIF needs to focus 1.8 megajoules of ultraviolet light (with a wavelength of 351 nanometers) into a volume with a diameter of 600 micrometers, and all beams are required to have a power output matching within 8% during the entire 20 - nanosecond waveform. To achieve this goal, the NIF's beam diagnostic system must have an unprecedented level of precision and automation. ### Main Problems and Challenges 1. **Power Output Consistency**: - It is necessary to ensure that the power output of all beams remains within an 8% deviation range during the entire 20 - nanosecond waveform. - The measurement accuracy of the total laser pulse energy needs to reach 2.8%, and the measurement accuracy of the temporal variation of the pulse power needs to reach 4%. 2. **Beam Path Control and Characterization**: - Each beam path contains 110 major optical components, distributed over a 510 - meter path. - The tolerance requirements of the diagnostic system are very strict. For example, the measurement accuracies of the temporal shapes of beam energy and power are 2.8% and a rise time of ≤450 picoseconds respectively. 3. **Spatial Distribution Measurement**: - It is necessary to record the spatial distribution of beam flux with a resolution of 2% and a spatial resolution of 1/125 of the beam. ### Solutions To solve the above problems, the paper proposes the following solutions: - **Sensor Packages and Diagnostic Sensors**: Use approximately 160 sensor packages and set up multiple measurement points on each beam path, including energy measurement, power measurement, and spatial distribution measurement. - **Automation and Computer Control Systems**: Through the Integrated Computer Control System (ICCS), highly automated operation of widely distributed sensors is achieved to ensure repeatability and accuracy. - **Data Processing and Calibration**: Use the Front - End Processor (FEP) and diagnostic electronics to capture and process data, and interface through the LonTalk fieldbus. In addition, an in - situ calibration system is also designed to reduce measurement errors. ### Conclusion The paper shows how, by combining commercial off - the - shelf technology (COTS) and the distributed system developed by Lawrence Livermore National Laboratory (LLNL), combined with in - situ calibration and digital correction techniques, high - precision measurement and control of the NIF beam diagnostic system are achieved. These systems ensure that the NIF can operate stably under strict performance requirements, thus supporting the achievement of its scientific experiment goals.